KEY FEATURES
|
|
|
INCUBATION OF DIPLIDE TESTS Dipslides are used to establish and asses if the conditions present in a water system encourage bacteria to multiply. Once the sample has been collected the dipslide requires incubation at 30°C for 48 hours to obtain a true result and consistently time after time.
A dipslide incubator is a critical element of microbiological monitoring. A lower temperature can mean the bacteria will grow more slowly and give a false low reading. If your reading is artificially low then the treatment engineer will consider that no action is necessary and the bacteria will be free to multiply and increase the risk to public health. Conversely if the temperature is high then the growth may be accelerated giving a false positive. This false high bacteria count can lead to unnecessary and expensive extra cleaning and treatment processes.
Approved code of practice and guidance for the control of legionella bacteria in water systems requires weekly monitoring of microbiological activity.
Gnat INCUBATOR The Gnat Incubator has a capacity to incubate two dipslides at a preset temperature of 30°C. The two heated dipslide cells will hold standard MicroSlide Dipslides or non-screw top dipslides.
Running on the supplied 220/240V mains power lead or an optional 12V power lead for use in a standard car lighter socket, it is protected by an internal fuse and has an over-temp protection for safe extended use.
|
|